NPi-300 Pupillometer

The NPi-300 Pupillometer is a reliable handheld device for objective measurement of pupil reactivity in intensive care and also has an integrated barcode reader for 1D and 2D codes.

  • Technical data:
    Infrared camera with light stimulus provided by LED lamps. Contactless inductive charging of the Li-ion battery.
  • MDD classification:
    Class 1 LED product in accordance with IEC 60825.
  • Measurement:

    During the automated measurement, a total of 90 images are captured within 3 seconds. Measurement accuracy: +/- 0.03 mm.

    The measurement result is displayed on the screen and saved on the SmartGuard. The data can be read out with a SmartGuard Reader and saved in a table as preparation for the PDMS.

  • Measured parameters:
    • NPi:
      Neurological Pupil index
    • MAX:
      Pupil diameter before light stimulus in mm
    • MIN:
      Smallest measured pupil diameter in mm
    • % CH:
      Percentage change in pupil size in %
    • CV:
      Average constriction velocity in mm/s
    • MCV:
      Maximum constriction velocity in mm/s
    • LAT:
      Time delay until reaction to light stimulus occurs in s
    • DV:
      Dilation velocity in mm/s
  • Evaluation of the NPi
    • 3,0 – 4,9:
      Normal pupillary response
    • < 3,0:
      Abnormal pupillary response*
    • 0:
      Non-reactive or atypical pupillary response

    *A difference in NPi between the right and left pupils of ≥ 0.7 may also be considered an abnormal pupil reading.

    The NPi is a numerical indicator of pupil reactivity. Any change in the NPi can be objectively determined, displayed as a trend, and documented for further action.